九一国产精品_日本免费不卡一区二区_在线a毛片_久久在草_国产一区二区影视_亚洲国产美女视频

技術中�

CSB-E09632h人游離血紅蛋白(f-Hb)ELISA試劑盒說明書

2011�12�20� 15:35:08人氣�324來源�廈門慧嘉生物科技有限公司

資料類型doc文件資料大小110787
下載次數108資料圖片 【點擊查看�
� � �廈門慧嘉生物科技有限公司 需要積�0
� � �人游離血紅蛋�,f-Hb,ELISA試劑�
【資料簡介�

 

Human free haemoglobin(f-Hb� ELISA Kit
Catalog No. CSB-E09632h
�96T�
This immunoassay kit allows for the in vitro quantitative determination of human f-Hb concentrations in serum, plasma and other biological fluids.
Expiration date six months from the date of manufacture
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
 
廈門慧嘉生物長期經營ELISA試劑�及Santa/Abcam抗體、Peprotech細胞因子、Sigma/Amresco/Qiagen、Axygen耗材等生物試劑產品。誠信經營,價格實惠,服務周到,質量有保證。歡迎廣告老師來詢�:  :  1048735792 或登�/download(向客服人員索取原版說明書)
 
PRINCIPLE OF THE ASSAY
The microtiter plate provided in this kit has been pre-coated with an antibody specific to f-Hb. Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated polyclonal antibody preparation specific for f-Hb. and Avidin conjugated to Horseradish Peroxidase (HRP� is added to each microplate well and incubated. Then a TMB �3,3'5, 5' tetramethyl-benzidine� substrate solution is added to each well. Only those wells that contain f-Hb, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of a sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm ±2 nm. The concentration of f-Hb. in the samples is then determined by comparing the O.D. of the samples to the standard curve.
DETECTION RANGE
1.56 μg/ml-100 μg/ml. The standard curve concentrations used for the ELISA’s were 100 μg/ml,50 μg/ml, 25 μg/ml, 12.5 μg/ml, 6.25 μg/ml, 3.12 μg/ml, 1.56 μg/ml.
SPECIFICITY
This assay recognizes recombinant and natural human f-Hb No significant cross-reactivity or interference was observed.
SENSITIVITY
The minimum detectable dose of human f-Hb is typically less than 0.39 μg/ml. The sensitivity of this assay, or Lower Limit of Detection (LLD� was defined as the lowest protein concentration that could be differentiated from zero.
MATERIALS PROVIDED

Reagent
Quantity
Assay plate
1
Standard
2
Sample Diluent
1 x 20 ml
Biotin-antibody Diluent
1 x 10 ml
HRP-avidin Diluent
1 x 10 ml
Biotin-antibody
1 x 120μl
HRP-avidin
1 x 120μl
 
1 x 20 ml
Wash Buffer
 
 
�25×concentrate�
TMB Substrate
1 x 10 ml
Stop Solution
1 x 10 ml

STORAGE
1          Unopened test kits should be stored at 2-8?C upon receipt and the microtiter plate should be kept in a sealed bag. The test kit may be used throughout the expiration date of the kit, provided it is stored as prescribed above. Refer to the package label for the expiration date.
2          Opened test plate should be stored at 2-8?C in the aluminum foil bag with desiccants to minimize exposure to damp air. The kits will remain stable until the expiring date shown, provided it is stored as prescribed above.
3          A microtiter plate reader with a bandwidth of 10 nm or less and an optical density range of 0-3 OD or greater at 450nm wavelength is acceptable for use in absorbance measurement.
 
REAGENT PREPARATION
Bring all reagents to room temperature before use.
1          Wash Buffer If crystals have formed in the concentrate, warm up to room temperature and mix gently until the crystals have compley dissolved. Dilute 20 ml of Wash Buffer Concentrate into deionized or distilled water to prepare 500 ml of Wash Buffer.
2          Standard Centrifuge the standard vial at 6000-10000rpm for 30s. Reconstitute the Standard with 1.0 ml of Sample Diluent. This reconstitution produces a stock solution of 100 μg/ml. Allow the standard to sit for a minimum of 15 minutes with gentle agitation prior to making serial dilutions. The undiluted standard serves as the high standard �100 μg/ml�. The Sample Diluent serves as the zero standard �0 μg/ml�. Prepare fresh for each assay. Use within 4 hours and discard after use.
3          Biotin-antibody Centrifuge the vial before opening. Dilute to the working concentration using Biotin-antibody Diluent�1:100�, respectively.
4          HRP-avidin Centrifuge the vial before opening. Dilute to the working concentration using HRP-avidin Diluent�1:100�, respectively.
 
Precaution: The Stop Solution provided with this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
OTHER SUPPLIES REQUIRED
1           Microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
2           Pipettes and pipette tips.
3           Deionized or distilled water.
4           Squirt bottle, manifold dispenser, or automated microplate washer.
5           An incubator which can provide stable incubation conditions up to 37°C±0.5°C.
 
SAMPLE COLLECTION AND STORAGE
Serum Use a serum separator tube (SST� and allow samples to clot for 30 minutes before centrifugation for 15 minutes at 1000 g. Remove serum and assay immediay or aliquot and store samples at -20°C. Centrifuge the sample again after thawing before the assay. Avoid repeated freeze-thaw cycles.
Plasma Collect plasma using citrate, EDTA, or heparin as an anticoagulant. Centrifuge for 15 minutes at 1000 g within 30 minutes of collection. Assay immediay or aliquot and store samples at -20°C. Centrifuge the sample again after thawing before the assay. Avoid repeated freeze-thaw cycles.
 
Note: Grossly hemolyzed samples are not suitable for use in this assay.
ASSAY PROCEDURE
Bring all reagents and samples to room temperature before use. It is recommended that all samples, standards, and controls be assayed in duplicate. All the reagents should be added directly to the liquid level in the well. The pipette should avoid contacting the inner wall of the well.
1          Add 100μl of Standard, Blank, or Sample per well. Cover with the adhesive strip. Incubate for 2 hours at 37°C.
2          Remove the liquid of each well, don’t wash.
3          Add 100μl of Biotin-antibody working solution to each well. Incubate for 1 hour at 37°C. Biotin-antibody working solution may appear cloudy. Warm up to room temperature and mix gently until solution appears uniform.
4          Aspirate each well and wash, repeating the process three times for a total of three washes. Wash: Fill each well with Wash Buffer �200μl� and let it stand for 2 minutes, then remove the liquid by flicking the plate over a sink. The remaining drops are removed by patting the plate on a paper towel. Complete removal of liquid at each step is essential to good performance.
5          Add 100μl of HRP-avidin working solution to each well. Cover the microtiter plate with a new adhesive strip. Incubate for 1 hour at 37°C.
6          Repeat the aspiration and wash three times as step 4.
7          Add 90μl of TMB Substrate to each well. Incubate for 10-30 minutes at 37°C. Keeping the plate away from drafts and other temperature fluctuations in the dark.
8          Add 50μl of Stop Solution to each well when the first four wells containing the highest concentration of standards develop obvious blue color. If color change does not appear uniform, gently tap the plate to ensure thorough mixing.
9          Determine the optical density of each well within 30 minutes, using a microplate reader set to 450 nm.
 
CALCULATION OF RESULTS
Using the professional soft "Curve Exert 1.3" to make a standard curve is recommended, which can be downloaded from our web.
Average the duplicate readings for each standard, control, and sample and subtract the average zero standard optical density. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic �4-PL� curve-fit. As an alternative, construct a standard curve by plotting the mean absorbance for each standard on the y-axis against the concentration on the x-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the f-Hb concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.
LIMITATIONS OF THE PROCEDURE
1           The kit should not be used beyond the expiration date on the kit label.
2           Do not mix or substitute reagents with those from other lots or sources.
3           It is important that the Standard Diluent selected for the standard curve be consistent with the samples being assayed.
4           If samples generate values higher than the highest standard, dilute the samples with the appropriate Standard Diluent and repeat the assay.
5           Any variation in Standard Diluent, operator, pipetting technique, washing technique, incubation time or temperature, and kit age can cause variation in binding.
6           This assay is designed to eliminate interference by soluble receptors, binding proteins, and other factors present in biological samples. Until all factors have been tested in the Quantikine Immunoassay, the possibility of interference cannot be excluded.
 
TECHNICAL HINTS
? Centrifuge vials before opening to collect contents.
 
1           When mixing or reconstituting protein solutions, always avoid foaming.
2           To avoid cross-contamination, change pipette tips between additions of each standard level, between sample additions, and between reagent additions. Also, use separate reservoirs for each reagent.
3           When using an automated plate washer, adding a 30 second soak period following the addition of wash buffer, and/or rotating the plate 180 degrees between wash steps may improve assay precision.
4           To ensure accurate results, proper adhesion of plate sealers during incubation steps is necessary.
5           Substrate Solution should remain colorless or light blue until added to the plate. Keep Substrate Solution protected from light. Substrate Solution should change from colorless or light blue to gradations of blue.
6           Stop Solution should be added to the plate in the same order as the Substrate Solution. The color developed in the wells will turn from blue to yellow upon addition of the Stop Solution. Wells that are green in color indicate that the Stop Solution has not mixed thoroughly with the Substrate Solution.

廈門慧嘉生物科技有限公司作�

上一篇:供應0.1� 0.3� 0.5級標準智能測力儀_可定制標準智能測力儀

下一篇:數字測力計規格_數字測力儀廠家_數字推拉力儀價格


我要投稿
  • 投稿請發送郵件至�(郵件標題請備注“投稿”)[email protected]
  • 聯系電話0571-87759680
環保行業“互聯網+”服務平�
環保在線APP

功能豐富 實時交流

環保在線小程�

訂閱獲取更多服務

微信公眾�

關注我�

抖音

環保在線�

抖音號:hbzhan

打開抖音 搜索頁掃一�

視頻�

環保在線

公眾號:環保在線

打開微信掃碼關注視頻�

快手

環保在線

快手ID�2537047074

打開快手 掃一掃關�
意見反饋
主站蜘蛛池模板: 亚洲成人综合网站 | 国产精品久久久久久久午夜片 | 草碰人人 | 国产精品剧情一区二区三区 | wankzhd| 全黄裸片武则天一级第4季 偿还电影免费看 | 一级做a爱视频 | 黄网站在线免费 | 9999视频| 天天夜干 | 国产精品亚洲精品日韩已方 | 一本色道久久综合狠狠躁篇适合什么人看 | 成人在线观看免费视频 | 717影院理论午夜伦八戒秦先生 | 91国内精品久久久久免费影院 | 国产亚洲精品久久久久久久久 | 国产午夜免费视频 | 亚洲午夜网站 | 亚洲成人伊人 | 亚州综合 | 国产精品久久国产精麻豆96堂 | 国产精品视频成人 | 国产一区国产二区在线观看 | 欧美一区二区三区中文字幕 | 国产成人观看 | 91久久一区 | 欧美一级黄色免费看 | 亚洲国产美女视频 | 91成人免费看 | 成人情欲视频在线看免费 | 97黄色网| 伦理三区 | 一级国产精品一级国产精品片 | 天天色图片 | 黄色视频一级毛片 | 少妇的肉体的满足毛片 | 九九热精品在线 | 99ri精品 | 色a综合| 午夜视频在线观看免费视频 | 久久在线 |